OEM Package CCD Spectrometer
                 

• Best performance cost ratio in the industry • Designed from the ground up for OEM integration • Small impact resistant optical bench • Flexible light input direct to slit or via fiber • Fully customizable • Sony ILX511 CCD for SM200 or Toshiba TCD1304 CCD for SM400 allows up to a 900 nm measurement window between 200nm and 1050nm (800 nm measurement window size possible for some applications) • USB 1.1 and 2.0 interface with 16-bit dynamic range available • Support up to 8 USB multi-channel configuration • UV enhancing coating

FEATUER VALUE
Model SM200 SM400
Detectors Sony ILX511(UV Enhanced Coated)
▶ Number of Pixels : 2048
▶ Sensing Pixel Size : 14 μm X 200 μm
▶ Sensitivity : 200V/(lx s) under 3200K
uniform illumination
Toshiba TCD 1304
▶ Number of Pixels : 3648
▶ Sensing Pixel Size : 8 μm X 200 μm
▶ Sensitivity : 160V/(lx s) under daylight
fluorescent illumination
Spectrograph # 2.7
Dark Noise RMS <50 RMS counts in 16 bits
SNR 250 : 1
Fiber Coupler SMA905 or FC standard
Wavelength Range 200 - 1050 nm
Order Sorting Filter Long pass filter or linear variable filter installed per wavelength coverage
Spectral Resolution 0.3 ~ 10 nm FWHM
Stray Light Level <0.05% @ 632 nm(<0.1% overall)
Computer Interface USB 1.1/2.0 compatible
Min. Integration Time 1 msec
Operating System Windows 7/10 (32/64 bit)
SDK Support Visual C++ / Visual Basic / LabVIEW
Dimensions / Weight 97mm X 68mm X 23mm / 0.2Kg

The Prime Choice for Integration with OEM Products

The SM200 is a miniature spectrometer designed for custom OEM applications. It offers a high performance to cost ratio for new systems designs. It can accept light directly through its built-in slit or via optical fiber. A removable fiber coupler faceplate allows use of standard SMA 905, FC and custom fiber connectors. This faceplate also allows direct attachment to dedicated systems and a number of SMX Accessories. A durable aluminum housing encloses the SM200 optical bench; through careful design this housing provides stable device operation over a wide range of temperatures.

The spectrometer sensor array and array driver electronics are mounted inside the SM200 housing, from there a flex cable connects to exterior support electronics. The standard sensor array used is the Sony ILX511. The driver electronics have been designed for highly sensitive yet stable operation. The design of the SM200 also allows the use of custom arrays for special applications, including photodiode assemblies and alternative CCD arrays, as well as the use of customer's own data acquisition electronics instead of using KSP's.

Standard interfaces to the SM200 include a USB 2.0 interface with 16-bit extended dynamic range and a PCI card interface with 16-bit dynamic range sampling. Our USB board can support multichannel configuration up to 8. With this multichannel configuration, a high resolution for wide range or a dual spectrometer system (one for measurement and the other for reference) is possible.

Spectral Products applies new UV enhancing coating on the CCD to increase the UV sensitivity below 450nm comparing with the conventional UV coating that is widely used in CCD spectrometers. By the help of this new UV coating, the signal sensitivity below 500nm can get improved ~20-50% more in general.

Software support includes custom DLLs for dedicated applications development and our SM32Pro windows based spectral acquisition and analysis software. Both standard and legacy interface designs provide support for advanced acquisition programming and external triggering.

ㅣ Software ㅣ Manual ㅣ Catalog ㅣ ▶  Download